Hi,

Alan Coelho helped us get our instrument parameter file set up for Topas Academic. We use an X'celerator detector with cobalt radiation (no monochromator) for most of our work.

Here is an example file:

xdd "test.dat"
        CoKa7_Holzer(0.001)
        Radius(240)
        LP_Factor(17)
        axial_conv
                filament_length 12
                sample_length 15
                receiving_slit_length 12
                primary_soller_angle @  5

Hope this helps.

Laurel Basciano


At 06:48 AM 5/2/2005, you wrote:
Dear colleagues,

Is there anybody who has experience with describing peak shapes
using the fundamental parameters approach for a typical X'pert-pro
diffractometer setup with an X'celerator detector (fixed incident
slits, 0.02 rad sollers, no monochromator)?

Thanks in advance, Arie

*******************************************
A. van der Lee
Institut Européen des Membranes (UMR 5635)
Université de Montpellier II - cc 047
Place E. Bataillon
34095 Montpellier Cedex 5
FRANCE

visiting address: 300 Av. Prof. E. Jeanbrau

tél.: 00-33-(0)4.67.14.91.35
FAX.: 00-33-(0)4.67.14.91.19
*******************************************

--
Laurel Basciano
Department of Geological Sciences and Geological Engineering
Queen's University
[EMAIL PROTECTED]
(613)484-1911






Reply via email to