Dear Aidong,

As Dr. Förster has rightly pointed out you need to have the latest version of 
HKL2000 to work with the HDF5 files from the Eiger detectors and as you must be 
aware the HKL2000 license should include use of Eiger detectors. We have 
recently used HKL2000 to process data from MASSIF-3 beamline at ESRF which uses 
a EIGER X 4M detector but we had to upgrade to the latest version of HKL2000.

With regards,
Raghu


From: CCP4 bulletin board [mailto:CCP4BB@JISCMAIL.AC.UK] On Behalf Of Andreas 
Förster
Sent: Tuesday, December 19, 2017 11:47
To: CCP4BB@JISCMAIL.AC.UK<mailto:CCP4BB@JISCMAIL.AC.UK>
Subject: Re: [ccp4bb] Eiger 16M detector in HKL2000

Dear Aidong,

as I don't have access to HKL2000 myself, I can't tell you what might need to 
be done to have the software display EIGER X 16M images.  Wladek showed me that 
it was possible.  It might just be a question of getting the latest version.

There are alternatives to HKL-2000.

DIALS can process EIGER data with no questions asked 
(https://dials.github.io/).  It is free, open-source software.  DIALS exists as 
a powerful command line tool but can also be run through the xia2 pipeline 
(https://xia2.github.io/).  Both are part of CCP4, but I'd recommend installing 
the latest DIALS nightly because development is rapid.  The developers respond 
quickly to questions.

XDS can process EIGER data 
(http://xds.mpimf-heidelberg.mpg.de/html_doc/XDS.html).  It is free to academic 
users.  To process data in HDF5 format (*.h5), make sure to download the neggia 
plugin (https://www.dectris.com/support/downloads/software/neggia, free, but 
registration required) and adjust the LIB= parameter to point to where the 
plugin is saved.  XDS is a command line tool, but several GUIs exist.

Please give DIALS or XDS a try and tell me how you get along.  If you continue 
to encounter problems, let me and the beamline staff at SSRF know.

All best.



Andreas



On Tue, Dec 19, 2017 at 6:48 AM, Aidong Han 
<a...@xmu.edu.cn<mailto:a...@xmu.edu.cn>> wrote:
Hi All,

We have recently collected several data sets in the Shanghai synchrotron 
facility with a newly installed Eiger 16M. However, we are not able to process 
these data using our home HKL2000 package. Even though we can easily add this 
detector in, we can not display images. I believe ours has not been properly 
set up. While we are still waiting for a response from HKL2000 administration 
managers (it has been a few days), I wonder whether someone can provide me your 
solution since this detector has been in market for more than a year. Thank you 
for your help.

Sincerely,

Aidong




Reply via email to