Hi Alex
(1) Go to “Settings -> Experiment settings -> Detector” and change the value
there; you should probably re-run integration, though since your value is only
just outside the range of 0.7 - 1.3, I might not worry too much about that, and
I don’t think it would make a big difference.
(2)
Dear CCP4bb members,
I am new to iMosflm data processing (I use HKL2000 but I really want to
learn iMosflm). After automatic index, cell refinement and integration in
iMosflm, I got 2 warnings.
1. Error in detector gain. BGRATIO of my data lies outside the 0.7 to 1.3
range (mine is 0.64) Imosfl