We have the BMP set up as the "priority panel" on our Dimension XPLUS analyzer. It works great, except if there happens to be a critical result on the panel, and there are remaining tests to result on a second transmission after the BMP is verified. When the remaining tests are sent to the analyzer batch screen, the critical value text is invoked *again* on the previously verified result, and Meditech is seeing the result as an edited verified result! Has anyone else seen this? What was done to resolve it? We really like the functionality of the priority panel, and want to keep it, but this is messy and annoying. We are Magic 5.5 SR3.
Thanks for any suggestions! Britt Watters, MT(ASCP) Laboratory System Analyst Bartlett Regional Hospital
=*=*=*=*=*=*=*=*=*=*=*=*=*=*=*=*=*=*=*= To subscribe or unsubscribe to the meditech-l, visit http://mtusers.com/mailman/listinfo/meditech-l_mtusers.com To check the status of the meditech-l, visit MTUsers.NET For help, email [EMAIL PROTECTED] Please visit and add information to the MTUsers WikiPedia at MTUsers.NET/mwiki ______________________________________ meditech-l mailing list meditech-l@MTUsers.com http://mtusers.com/mailman/listinfo/meditech-l_mtusers.com