Dear All, As far as I know it could happen when you are using i2. We are working to fix this problem. At the moment the best solution is to use advanced options in the refmac interface and define resolution limits explicitly. For example you can add in the advanced options:
resolution 37 2.5 where these resolution limits are from the aimless output. Low resolution limit should not affect refinement behaviour, however it may cause problem during deposition. We will have better solution soon. I2 has many good features. Please report if you see some misbehaviours. Best wishes, Garib P.S. you can also use advanced options to do things that are not available on the interface yet. 1) refinement against electron diffraction. Add source EC MB # electron form factor will be used using Mott-Bethe formula 2) you can add occupancy refinement etc On 18 Mar 2018, at 10:46, "Orru, Dr. Roberto" <ro...@uni-mainz.de> wrote: > Dear All, > > I am noticing that the low resolution in the reflections files after scaling > with aimless and after refinement with refmac does not coincide. > In a case I have 37A with aimless but for some reason refmac is 67A. > > Any idea? > All the best, > R. Dr Garib N Murshudov MRC-LMB Francis Crick Avenue Cambridge CB2 0QH UK Web http://www.mrc-lmb.cam.ac.uk, http://www2.mrc-lmb.cam.ac.uk/groups/murshudov/