Hi,

I would guess radiation damage, see Weik et al., 2000, PNAS 97, 623-628 or for 
a more recent discussion and thorough overview Sutton et al., 2013, Acta Cryst 
D69, 2381-2394.

best regards

Marjolein


On 02 Jun 2014, at 07:08, Eze Chivi 
<ezech...@outlook.com.ar<mailto:ezech...@outlook.com.ar>> wrote:

Hello, when I refine my structure, I see negative density around the disulfide 
bond. I have 7 copies per ASU, and I can see this density in many of them. In 
some cases, I see positive density also (negative in the center of the straight 
line linking S atoms, and positive in both sides). What can I try to solve it? 
Is it due to radiation damage? Alternative conformation (partial oxidation)? 
Incorrect disulfide geometry parameters? My resolution is 2.1 A, R/Rfree are 
around 0.220/0.243, and similar results with refmac5, phenix and PDBREDO. 
Please find two example pictures in attachment.
Thanks for your help!


Ezequiel
<disulfide.jpg><disulfide2.jpg>

[cid:3A2194AC-734C-45EB-B5A7-CE7FC0F0F1BC]




Marjolein Thunnissen
Science Coordinator MX

MAX IV Laboratory
Lund University
P.O. Box 118, SE-221 00 Lund, Sweden
Visiting address: Ole Römers väg 1, 223 63 Lund
Telephone: +46 766 32 04 17
www.maxlab.lu.se<http://www.maxlab.lu.se/>

Reply via email to