Using an envelope or a mask to start phasing even with high NCS is still worse 
than using a low res model from cryoEM that is characterized by an (almost) 
proper density distribution.

One has to measure the intensities of the very low res reflections (0 0 1 and 
the likes) precisely.

The presence of high NCS does not guarantee a success even if your initial 
model is "good" as described in Section 9 here:
http://www.ncbi.nlm.nih.gov/pubmed/11526317

Petr
------------------
Petr Leiman
EPFL
IPSB-LBBS
BSP-415
CH-1015 Lausanne
Suisse

Reply via email to