On Tue, Mar 24, 2015 at 11:07 PM, Arjun Nadh wrote:
> Thank you, Tom
> That fixes the problem.
Great. Thanks for testing.
Upcoming UHD and gr-uhd changes will expose more of the AD9361
functionality, which will allow user control of calibration and
filters without having to patch UHD. We're also
Hi Tom,
I have tried reducing the signal level. The signal level at the input is
now -40dBm. Still there is this ramp behaviour.
I also get this when I use the B210 in loopback mode (with a 30dB
atttenuator in the loop). Also I have observed that
the lower the master clock rate that I set, the le
Hi Arjun,
On Fri, Mar 20, 2015 at 11:15 PM, Arjun Nadh via USRP-users
wrote:
> On observing the received samples, there is a periodic " ramp " on the
> received sample. Attaching the plot. I do not observe this effect while
> using N210. Also I do not observe this if I use the internal clock.
Ca