Hi Sampath
That message should never appear during refinement; the popup is called
from the autoindexing code in iMosflm, and once that process is completed
successfully there is no way the code can call it - so there's obviously
something very odd happening. Indexing appears to have succeeded (i.
Have you picked from multiple images and do you have more than 1000 reflections
to index from?
You might also want to limit the resolution manually to perhaps 3 first.
Before processing your data you should refine your cell by using multiple
wedges of 10 degrees.
Check out the manual for additio
Dear All,
I have the problem in processing the data with imosflm. The data was
collected with 2.5A resolution about 300 images. Auto index found the space
group P222 with unit cell a= 40.2; b= 79.2; c= 194.7 and α=b=γ = 90.0. For
this cell, the penalty is 7. I feel this penalty is good enough to