Dear Simon, 

I don't know how to change the output written to the log-file (maybe with the
TEST-card), but I wonder why you don't want to look at all of your data but only
those with I/sigI > 0? XDS reports all reflections with I/sigI > -3 for a good
reason.

Cheers, Tim

On Tue, Feb 08, 2011 at 12:17:35PM +0000, Simon Kolstoe wrote:
> Dear ccp4bb,
> 
> I am quite a fan of XDS and have just upgraded to the latest version.
> 
> Normally, to assess the quality of my data, I look at the tables in
> CORRECT.LP and especially the table "SUBSET OF INTENSITY DATA WITH
> SIGNAL/NOISE >=  0.0 AS FUNCTION OF RESOLUTION".
> 
> However in my latest run I only get a single table for all the data
> i.e. for signal/noise >= -3.0. Is there a command I can put in my
> XDS.INP that will give me all the other tables or has the CORRECT.LP
> logfile been altered in the most recent version of XDS?
> 
> (FYI my xds.inp obtained from the ESRF last week is copied below)
> 
> Thanks,
> 
> Simon
> 
> 
> 
> !=== File Automaticaly generated by mxCuBE
>    !=== X-Ray data collected at: ESRF_ID14-1
>    !=== Detector type: ADSC Quantum Q210
>    !=== Date: Fri Feb 04 03:39:09 2011
>    !=== User comments:
> 
>    JOB= ALL !XYCORR INIT COLSPOT IDXREF DEFPIX XPLAN INTEGRATE CORRECT
>    !JOB= DEFPIX XPLAN INTEGRATE CORRECT
> 
>    DATA_RANGE= 1 190
>    SPOT_RANGE= 1 20
>    SPOT_RANGE= 1 4
>    !SPOT_RANGE= 187 190
>    BACKGROUND_RANGE= 1 4
> 
>    SECONDS=60
>    MINIMUM_NUMBER_OF_PIXELS_IN_A_SPOT= 6
>    STRONG_PIXEL= 6.0
> 
>    OSCILLATION_RANGE= 1.000
>    STARTING_ANGLE= 0.000
>    STARTING_FRAME= 1
>    X-RAY_WAVELENGTH=  0.93340
>    NAME_TEMPLATE_OF_DATA_FRAMES= ./data/sk1_1_???.img
> 
>    !STARTING_ANGLES_OF_SPINDLE_ROTATION= 0 180 10
>    !TOTAL_SPINDLE_ROTATION_RANGES= 60 180 10
> 
>    DETECTOR_DISTANCE= 298.55
>    DETECTOR= ADSC   MINIMUM_VALID_PIXEL_VALUE= 1   OVERLOAD= 65000
>    ORGX= 1014.79    ORGY= 1029.10
>    NX=  2048   NY=  2048   QX= 0.10200   QY= 0.10200
>    VALUE_RANGE_FOR_TRUSTED_DETECTOR_PIXELS= 7000 30000
> 
>    DIRECTION_OF_DETECTOR_X-AXIS= 1.0 0.0 0.0
>    DIRECTION_OF_DETECTOR_Y-AXIS= 0.0 1.0 0.0
>    ROTATION_AXIS= 1.0 0.0 0.0
>    INCIDENT_BEAM_DIRECTION= 0.0 0.0 1.0
>    FRACTION_OF_POLARIZATION= 0.98
>    POLARIZATION_PLANE_NORMAL= 0.0 1.0 0.0
>    !== Default value recommended
>    !AIR= 0.00026895
> 
>    SPACE_GROUP_NUMBER= 0
>    UNIT_CELL_CONSTANTS= 0 0 0 0 0 0
>    INCLUDE_RESOLUTION_RANGE= 50.0 2.4
>    RESOLUTION_SHELLS= 15.0 8.0 4.0 3.0 2.8 2.6 2.5 2.4
>    FRIEDEL'S_LAW= FALSE
> 
>    !FRIEDEL'S_LAW= TRUE
>    TRUSTED_REGION= 0 1.40
> 
>    REFINE(INTEGRATE)= BEAM ORIENTATION CELL
>    !== Default value recommended
>    !DELPHI= 3.000
>    MAXIMUM_NUMBER_OF_PROCESSORS= 16
>    !MAXIMUM_NUMBER_OF_JOBS= 16

-- 
--
Tim Gruene
Institut fuer anorganische Chemie
Tammannstr. 4
D-37077 Goettingen

phone: +49 (0)551 39 22149

GPG Key ID = A46BEE1A

Attachment: signature.asc
Description: Digital signature

Reply via email to